Valivera

Tools and resources

Practical tools and references for medical device teams, from shelf-life calculations to the standards and FDA documents we work with every day.

FDA device and pathway checker

Answer a few questions to see whether FDA would likely treat your product as a medical device, which class it falls in, and which premarket submission it would need.

  1. Is it a device?
  2. Which class?
  3. Which submission?

Step 1 of 3 · Is it a device?

What is your product intended to do?

FDA looks at intended use: what you claim the product does, in labeling, marketing and on your website.

For education only, not a regulatory determination. Classification depends on your exact intended use and product code, and FDA makes the final call.

Accelerated aging time calculator

Estimate how long samples need in the chamber to simulate a target shelf life, using the Arrhenius-based method in ASTM F1980.

Accelerated aging time

46 days

In weeks

6.5

Aging factor (AAF)

8.00

AAF = Q10(T_AA − T_RT)/10, AAT = real time ÷ AAF, with one month taken as 30.44 days. An estimate for planning; confirm it against ASTM F1980 and your test protocol.

French to millimeters converter

Convert catheter and sheath sizes between French, millimeters and inches.

Type in any box. French size is the outer diameter in thirds of a millimeter: 1 Fr = 0.333 mm, so 9 Fr = 3 mm.